| Product Name | Glucose LiquiuUV Mono Test (Hexokinase Method) |
| Catalog No. | BSTR-0002 |
| Description | This glucose test uses the hexokinase method, which is highly specific and sensitive for glucose detection. As a single reagent, it simplifies operation and reduces experimental errors, suitable for precise quantitative analysis in various biological fluid samples for scientific research. |
| Test Target | Glucose (Glu) |
| Methodology | Enzymatic colorimetric (Hexokinase, Glucose-6-Phosphate Dehydrogenase) |
| Measurement Range | 0.3-25 mmol/L |
| Sample Type | Serum, plasma, cerebrospinal fluid, urine (diluted) |
| Assay Time | 10 min at 37°C |
| Wavelength | 340 nm |
| Linearity | 0.3-25 mmol/L (R²≥0.998) |
| Precision (Intra-assay CV) | ≤1.5% |
| Precision (Inter-assay CV) | ≤2.5% |
| Stability (Unopened) | 24 months at 2-8°C |
| Stability (Opened) | 45 days at 2-8°C |
| Stability (Reconstituted) | 14 days at 2-8°C |
| Interference (Hemoglobin) | ≤600 mg/dL (no significant interference) |
| Interference (Ascorbic Acid) | ≤100 mg/dL (no significant interference) |
| Interference (Uric Acid) | ≤150 mg/dL (no significant interference) |
| Reagent Composition | Hexokinase, glucose-6-phosphate dehydrogenase, ATP, NADP+, PIPES buffer, magnesium ions |
| Reagent Form | Liquid (single reagent) |
| Reagent Volume | 4×100mL; 1×3mL (standard) |
| Calibration Type | External calibration |
| Calibration Levels | 4 levels (0, 5.55, 11.1, 22.2 mmol/L) |
| Incubation Temperature | 37°C |
| Detection Instrument | Spectrophotometer, clinical chemistry analyzer |
| Application | Quantitative determination of glucose in biological fluids for research on glucose metabolism disorders, neurological diseases, and renal glucose excretion functions |
| Cross - Reactivity | No cross-reactivity with common monosaccharides and disaccharides |
| pH Range | 7.4-7.8 |
| Shelf Life | 24 months |
| Storage Condition | Store at 2-8°C, avoid freezing and light, seal tightly after opening |
| Minimum Detection Limit | 0.1 mmol/L |
| Assay Format | Test tube / cuvette |