| Product Name | Blood Glucose Sensors (Dry Chemical Method) |
| Catalog No. | BSTR-0004 |
| Description | This blood glucose sensor uses dry chemical technology for rapid glucose detection. It is single-use, easy to operate, and suitable for on-site research and rapid sample screening, providing reliable results for preliminary glucose metabolism studies. |
| Test Target | Glucose (Glu) |
| Methodology | Dry chemical (Glucose Oxidase, Electron Mediator) |
| Measurement Range | 0.5-33.3 mmol/L |
| Sample Type | Whole blood (capillary, venous), serum, plasma |
| Assay Time | 5 min at room temperature (20-25°C) |
| Wavelength | Reflectance photometry (550-600 nm) |
| Linearity | 0.5-33.3 mmol/L (R²≥0.995) |
| Precision (Intra-assay CV) | ≤3% |
| Precision (Inter-assay CV) | ≤4% |
| Stability (Unopened) | 18 months at 2-30°C |
| Stability (Opened) | N/A (single-use) |
| Stability (Reconstituted) | N/A (single-use) |
| Interference (Hemoglobin) | ≤500 mg/dL (no significant interference) |
| Interference (Ascorbic Acid) | ≤50 mg/dL (no significant interference) |
| Interference (Uric Acid) | ≤100 mg/dL (no significant interference) |
| Reagent Composition | Glucose oxidase (Aspergillus niger), electron mediator, buffer, stabilizer, desiccant |
| Reagent Form | Dry strip |
| Reagent Volume | 25 strips/box, 50 strips/box |
| Calibration Type | Built-in calibration |
| Calibration Levels | 2 levels (5.55, 11.1 mmol/L) |
| Incubation Temperature | Room temperature (20-25°C) |
| Detection Instrument | Dry biochemical analyzer, blood glucose meter |
| Application | Rapid quantitative determination of glucose in whole blood/serum/plasma for on-site research, rapid sample screening, and glucose metabolism preliminary studies |
| Cross - Reactivity | No cross-reactivity with maltose, galactose, xylose |
| pH Range | 7.0-7.5 |
| Shelf Life | 18 months |
| Storage Condition | Store at 2-30°C, dry place, avoid light and moisture, seal tightly after opening |
| Minimum Detection Limit | 0.2 mmol/L |
| Assay Format | Single-use dry strip |
For research use only, not for clinical use.
|