Research
Online Inquiry

Very Low - Density Lipoprotein Cholesterol (VLDL - C) Assay Kit

Cat.No: BLTR-0009 Datasheet

Quantities:
- +
Product Details Related Products
Product Name Very Low - Density Lipoprotein Cholesterol (VLDL - C) Assay Kit
Catalog No. BLTR-0009
Description This Very Low - Density Lipoprotein Cholesterol assay kit uses a selective enzymatic reaction method for direct quantitative detection of VLDL - C in serum and plasma. It has good specificity, acceptable precision, and a suitable linear range, supporting research on VLDL - C - related lipid metabolism disorders
Test Target Very Low - Density Lipoprotein Cholesterol (VLDL - C)
Methodology Enzymatic colorimetric (selective enzymatic reaction)
Measurement Range 0.2-2.0 mmol/L
Sample Type Serum, plasma
Assay Time 15 min at 37°C
Wavelength 505 nm
Linearity 0.2-2.0 mmol/L
Precision (Intra-assay CV) ≤3%
Precision (Inter-assay CV) ≤4%
Stability (Unopened) 18 months at 2-8°C
Stability (Opened) 28 days at 2-8°C
Stability (Reconstituted) 10 days at 2-8°C
Interference (Hemoglobin) ≤500 mg/dL (no significant interference)
Interference (Bilirubin) ≤20 mg/dL (no significant interference)
Interference (Lipemic) ≤1000 mg/dL (no significant interference)
Reagent Composition Specific enzymes for VLDL - C, cholesterol oxidase, peroxidase, 4 - aminoantipyrine
Reagent Form Liquid
Reagent Volume R1: 45 mL×2; R2: 20 mL×2
Calibration Type External calibration
Calibration Levels 4 levels
Incubation Temperature 37°C
Detection Instrument Clinical chemistry analyzer
Application Quantitative determination of VLDL - C in serum/plasma to study its role in lipid metabolism and atherosclerosis
Cross - Reactivity No cross - reactivity with HDL, LDL
pH Range 7.0-7.5
Shelf Life 18 months
Storage Condition Store unopened at 2-8°C; opened reagents at 2-8°C

For research use only, not for clinical use.