Human Lyme Disease IgG Antibody ELISA Test Kit
Research
Online Inquiry

Human Lyme Disease IgG Antibody ELISA Test Kit

Cat.No: CS-DB-0044 Datasheet

Quantities:
- +
Contact Us for Pricing
Product Details Related Products
Product Name Human Lyme Disease IgG Antibody ELISA Test Kit
Catalog No. CS-DB-0044
Product Overview This ELISA test kit is designed for qualitative and quantitative detection of Lyme disease-associated IgG antibodies in human serum and plasma. It uses an enzyme-linked colorimetric reaction to evaluate IgG antibody reactivity, providing a method for laboratory investigation of serological responses associated with Borrelia infection.
Applications Qualitative assessment of Lyme disease-associated IgG antibody reactivity; quantitative serum and plasma antibody analysis; investigation of immune responses to Lyme-associated Borrelia species; laboratory serological evaluation in research and assay-development settings.
Detection Target Human Lyme disease IgG antibodies (Lyme IgG).
Sample Type Human serum and plasma
Result Type Qualitative and quantitative.
Pack Size 48T, 96T
Storage Conditions 2–8°C
Shelf Life 6 months
Methodology ELISA
Key Features Supports both qualitative and quantitative antibody analysis.
Uses a two-stage incubation process before color development.
Includes positive and negative controls for assay evaluation.
Supports optical density measurement at 450 nm with a 600–650 nm reference wavelength.
The source describes applicability to multiple Lyme-associated spirochete subtypes; this coverage requires verification.
Test Procedure Add 100 μL of standards, controls, or diluted samples to the appropriate coated microplate wells. For qualitative testing, the source specifies Standard B.
Cover the plate and incubate at 37°C for 60 minutes.
Remove the reaction solution and wash each well three times with 300 μL of prepared wash buffer. Remove residual liquid.
Add 100 μL of enzyme conjugate to each well.
Cover the plate and incubate at 37°C for 30 minutes.
Remove the reaction solution and wash each well three times with 300 μL of prepared wash buffer.
Add 100 μL of TMB substrate solution to each well.
Incubate at 18–25°C for 30 minutes, protected from light.
Add 100 μL of TMB stop solution to each well, maintaining consistent dispensing intervals.
Measure absorbance at 450 nm within 60 minutes, using a reference wavelength of 600–650 nm where applicable.
Result Interpretation For qualitative testing, evaluate sample reactivity against the designated standard or cutoff according to the assay instructions. For quantitative testing, determine antibody levels using the standards and calibration procedure. Exact qualitative thresholds and the validated measurement range have not been provided.
Package Contents Coated plate
Enzyme-linked conjugate
Standard
Positive control
Negative control
Dilution buffer
Wash solution
TMB substrate solution
TMB stop solution
Notes This test kit is intended for in vitro diagnostic use by professionals only.
Before beginning the test, carefully and thoroughly read the instructions to ensure that all procedures are clear.
For your safety, do not use any reagents that are damaged during the test.
Follow the lot numbers and expiration dates; do not mix reagents from different lots, and do not use expired reagents.
Adhere to good laboratory practices and safety guidelines. Wear a lab coat, rubber gloves, and safety goggles when necessary.
Avoid contact with the stop solution to prevent skin irritation or burns.
All reagents in the kit should be handled as potentially biohazardous materials during use and disposal.

For research use only, not for clinical use.