Anaplasma, Babesia and Ehrlichia Multiplex qPCR Kit
Research
Online Inquiry

Anaplasma, Babesia and Ehrlichia Multiplex qPCR Kit

Cat.No: CS-DB-0052 Datasheet

Quantities:
- +
Contact Us for Pricing
Product Details Related Products
Product Name Anaplasma, Babesia and Ehrlichia Multiplex qPCR Kit
Catalog No. CS-DB-0052
Product Overview A four-channel multiplex assay that screens three tick-borne pathogen groups in a single fluorescence PCR reaction while monitoring a separate internal DNA control. Its target set covers Anaplasma, Babesia and Ehrlichia sequences rather than Lyme-associated Borrelia.
Applications Multipathogen tick-borne research; evaluation of bacterial-protozoan coinfection patterns; laboratory assessment of multiplex assay workflow efficiency; epidemiologic monitoring of three target groups.
Detection Target Anaplasma spp. DNA, Babesia spp. (mammalian-host group) DNA and Ehrlichia spp. DNA.
Target Gene(s) Babesia spp.: 18S rRNA; Ehrlichia spp.: 16S rRNA; Anaplasma spp.: 23S; separate control marker.
Result Type Qualitative multiplex detection; concurrent positive channels may indicate more than one detectable pathogen group.
Reaction Volume 20 μL (11 μL primer/probe + master-mix premix; 9 μL input).
Pack Size 100T
Storage Conditions -20°C
Methodology PCR
Key Features Three pathogen groups analyzed in one well; four independent detection channels; Babesia-specific 18S / bacterial 16S and 23S targets; independent CY5 extraction control; coinfection-compatible interpretation.
Test Procedure Reconstitute dried positive and internal-control templates using the specified resuspension reagent.
Add 5 μL internal-control DNA into each sample extraction/lysis buffer, then extract DNA.
Prepare 11 μL reaction mixture per well (10 μL 2× qPCR master mix plus 1 μL multiplex primer/probe mix).
Add 9 μL extracted DNA, water (NTC) or positive-control template as appropriate.
Amplify at 95°C for 3 min, then 45 cycles of 95°C for 15 s and 60°C for 60 s, monitoring ROX, VIC/HEX, FAM and CY5.
Confirm target-specific and internal-control amplification before assigning results.

For research use only, not for clinical use.

0
0