| Product Name | One-Minute Swab Sample Release Reagent for Direct PCR and Fluorescent LAMP |
| Catalog No. | ATR-SPS-0093 |
| Description | Alta DiagnoTech sample release reagent is an ambient-stable liquid for direct molecular testing without conventional nucleic-acid extraction. |
| Intended Use | For rapid preparation of nasal or throat swabs for qPCR, RT-PCR, RT-LAMP or fluorescent LAMP. |
| Sample Type | Nasal swabs and throat swabs |
| Performance Range / Specifications | Batch consistency testing used one nasal swab and 1,000 copies of a pseudovirus target in 800 µL reagent; amplification curves were similar and ΔCt remained within ±0.5 Ct. |
| Reaction Conditions / Protocol | Place 800 µL reagent in a 1.5 mL collection tube. Insert the collected swab, rotate the swab head in the liquid for at least 30 seconds, and incubate for 1 minute at 20-25°C. A 25 µL RT-PCR or RT-LAMP reaction can accept up to 10 µL lysate. |
| Storage Conditions | Store at ambient temperature. |
| Package Specifications | Available in 10 mL, 100 mL and 1,000 mL formats. |
| Product Form | Liquid sample-release reagent |
| Shipping Conditions | Ship at ambient temperature. |
| Compatibility | Compatible with buffered PCR, RT-PCR and fluorescent LAMP systems. Not compatible with unbuffered colorimetric LAMP because the lysate is acidic. |
| Application Notes / Precautions | Reagent volume may be reduced to 400 µL per swab. Proceed to amplification as soon as possible after treatment, preferably within 3 hours. |
| Product Type | Direct-amplification sample lysis reagent |
| Function / Efficacy | Releases nucleic acid from swab samples for direct amplification and shortens sample preparation time. |