Biochemistry Analyzer Acid Wash Solution, 5% Phosphoric Acid-Based
Research
Online Inquiry

Biochemistry Analyzer Acid Wash Solution, 5% Phosphoric Acid-Based

Cat.No: ATRCS-0018 Datasheet

Quantities:
- +
Product Details Related Products
Product Name Biochemistry Analyzer Acid Wash Solution, 5% Phosphoric Acid-Based
Catalog No. ATRCS-0018
Description A phosphoric acid-based wash solution for removing inorganic scale deposits, calcium oxalate crystals, and mineral precipitates from biochemistry analyzer cuvettes, reagent probes, and fluid transfer lines. Formulated to dissolve deposits without attacking polycarbonate or quartz cuvette materials.
Intended Use Periodic descaling and maintenance cleaning of automated clinical chemistry analyzer cuvette wash stations and fluidic pathways.
Principle / Technology Controlled acid dissolution of mineral scale and inorganic precipitates at low pH
Detection Method Ready-to-use; no dilution required; aspirate through instrument fluidics for 5–10 minute contact time per manufacturer protocol
Components / Formulation Phosphoric acid (5% v/v), non-ionic wetting agent, corrosion inhibitor, deionized water
Storage Conditions Store at 15–25 °C in original tightly closed container; avoid contact with strong bases
Shelf Life 36 months from manufacture date
Package Specifications 500 mL, 2.5 L HDPE bottles
Product Form Clear colorless liquid
Key Features Optimized for polycarbonate cuvette safety; contains corrosion inhibitor for stainless steel components; phosphate-free alternative available for instruments with phosphate sensitivity; validated on Roche Cobas, Abbott Architect, Siemens ADVIA, and Beckman AU series analyzers
Purity Phosphoric acid 5.0 ± 0.2% (w/v); meets ACS reagent specifications
Concentration As specified (concentrate or ready-to-use)
Activity / Unit Definition Cleaning efficacy validated by instrument carryover testing
Molecular Weight Not applicable — detergent/surfactant formulations
Source / Origin Synthetic detergent and surfactant components
pH Range / Optimal pH pH 1.5–2.0 (neat solution at 25 °C)
Shipping Conditions Ambient temperature; no special handling required
Expiration Date / Stability 36 months under recommended storage; avoid prolonged contact with atmospheric CO2 which may cause buffer shift
Regulatory / Compliance CE-IVD marked; manufactured under ISO 13485; compliant with EU IVDR 2017/746 for instrument maintenance reagents
Compatibility Safe for polycarbonate, quartz glass, stainless steel 316L, PEEK, PTFE, and ETFE; avoid prolonged contact with aluminum, zinc, or galvanized surfaces
Recommended Buffer System Use undiluted or dilute per instrument manufacturer specification
Application Notes / Precautions Run instrument manufacturer's recommended acid wash cycle. After acid wash, perform minimum 3 rinse cycles with deionized water. Verify pH neutrality of final rinse waste before resuming patient testing. Recommend weekly preventive acid wash for high-throughput analyzers.
Batch-to-Batch Consistency pH and conductivity within ±5% of reference lot; cleaning efficacy validated

For research use only, not for clinical use.