- Home
- IVD
- By Technology Types
- By Diseases Types
- By Product Types
- Research
- Resource
- Distributors
- Company
| Product Name | Human Bocavirus 1-4 qPCR Detection Kit 200 copies/mL LOD |
| Catalog No. | MDKPK-0085 |
| Description | The Alta DiagnoTech human bocavirus nucleic acid detection kit uses real-time fluorescent PCR and covers all four currently identified HBoV genotypes. An internal control monitors the testing process, while UDG and dUTP help reduce carryover-related false-positive results. |
| Intended Use | Detection of human bocavirus genotypes 1-4 nucleic acids. |
| Principle / Technology | Real-time fluorescent PCR with internal process monitoring and UDG/dUTP contamination control. |
| Detection Method | Real-time fluorescent PCR |
| Performance Range / Specifications | Coverage of HBoV genotypes 1-4; stated detection limit is 200 copies/mL. |
| Sensitivity / LOD | 200 copies/mL |
| Components / Formulation | Internal control, UDG enzyme and dUTP contamination-control components. |
| Quality Control | Internal control monitors the laboratory testing workflow and helps reduce false-negative results. |
| Key Features | HBoV genotype 1-4 coverage 200 copies/mL detection limit Internal process control UDG/dUTP carryover prevention |
| Product Type | Respiratory viral nucleic acid detection kit |
| Analytes / Targets | Human bocavirus genotypes 1, 2, 3 and 4 |
| Genotypes / Variants Detected | Human bocavirus genotypes 1-4 |
| Clinical Agreement | Compared with Sanger sequencing, reported positive agreement was 100.00%, negative agreement was 98.38% and overall agreement was 98.70%. |
For research use only, not for clinical use.
|
There is no product in your cart. |