- Home
- IVD
- By Technology Types
- By Diseases Types
- By Product Types
- Research
- Resource
- Distributors
- Company
| Product Name | Urinary Transferrin Reagent |
| Catalog No. | KFTR-0015 |
| Description | A sensitive indicator for early glomerular basement membrane damage, often appearing alongside microalbumin. |
| Methodology | Immunoturbidimetry |
| Intended Use | Glomerular Permeability |
| Sample Type | Urine |
| Wavelength | 340 nm |
| Reaction Temp | 37°C |
| Reaction Type | Fixed-time |
| Direction | Increasing |
| Linearity Range | 0.5 - 100 mg/L |
| Sensitivity | ΔA/min ≥ 0.01 |
| Precision (Within-run) | CV < 4% |
| Precision (Between-run) | CV < 7% |
| Storage | 2-8°C |
| Stability (Opened) | 28 days |
| Blank Absorbance | ≤ 0.600 |
| Prozone Limit | 500 mg/L |
| Calibration | Multi-point |
| Bilirubin Interf. | 200 mg/L |
| Hemoglobin Interf. | 5 g/L |
| Lipid Interf. | 600 mg/dL |
| Ascorbic Acid | 50 mg/dL |
| R1/R2 Ratio | 0.16736111111111 |
| Reading Time | 5 Min |
| Unit | mg/L |
| Shelf Life | 12 Months |
For research use only, not for clinical use.
|
There is no product in your cart. |