| Product Name | Cystatin C (CysC) Kit |
| Catalog No. | KFTR-0004 |
| Description | Sensitive marker for GFR, independent of muscle mass or dietary intake for precise glomerular assessment. |
| Methodology | Latex Immunoturbidimetry |
| Intended Use | GFR change detection |
| Sample Type | Serum, Plasma |
| Wavelength | 570 nm |
| Reaction Temp | 37°C |
| Reaction Type | Fixed-time |
| Direction | Increasing |
| Linearity Range | 0.2 - 8.0 mg/L |
| Sensitivity | ΔA/min ≥ 0.05 |
| Precision (Within-run) | CV < 5% |
| Precision (Between-run) | CV < 6% |
| Storage | 2-8°C |
| Stability (Opened) | 28 days |
| Blank Absorbance | ≤ 0.800 |
| Prozone Limit | 30 mg/L |
| Calibration | Multi-point |
| Bilirubin Interf. | 300 mg/L |
| Hemoglobin Interf. | 5 g/L |
| Lipid Interf. | 500 mg/dL |
| Ascorbic Acid | 50 mg/dL |
| R1/R2 Ratio | 0.16736111111111 |
| Reading Time | 5 Min |
| Unit | mg/L |
| Shelf Life | 12 Months |