Lyme Disease IgG/IgM Lateral Flow Test Kit – 1/5/25 Tests
IVD
Online Inquiry

Lyme Disease IgG/IgM Lateral Flow Test Kit – 1/5/25 Tests

Cat.No: BI-QCY-0050 Datasheet

Quantities:
- +
Product Details Related Products
Product Name Lyme Disease IgG/IgM Lateral Flow Test Kit – 1/5/25 Tests
Catalog No. BI-QCY-0050
Product Overview A rapid qualitative immunoassay for detection of anti-Borrelia IgG and IgM antibodies. Suitable for preliminary screening; reactive results require interpretation alongside clinical findings and appropriate confirmatory testing.
Applications Professional screening for antibodies associated with Lyme disease; support for clinical assessment and subsequent confirmatory testing when indicated.
Certificate CE
Detection Target Anti-Borrelia IgG/IgM antibodies
Sample Type Whole blood, serum, and plasma samples
Result Type Qualitative (positive/negative)
Time to Result 10-15 minutes
Pack Size 1T, 5T, 25T
Storage Conditions 2–30°C
Sensitivity 93.96%(IgG), 90.50% (IgM)
Specificity 91.42%(IgG), 93.20%(IgM)
Methodology Lateral flow chromatographic immunoassay
Test Procedure Sampling. Collect human Serum/Plasma/Whole blood properly.
Testing. Remove an extraction tube from the kit and a test box from the film bag by tearing the notch. Put them on the horizontal plane. Open the inspection card aluminum foil bag. Remove the test card and place it horizontally on a table. Use a disposable pipette, transfer 4μL serum (or plasma), or 4μL whole blood into the sample well on the test cassette. Open the buffer tube by twisting off the top. Put 3 drops (about 80 μL) of assay diluent into the assay diluent well round shaped. Srart counting.
Reading. Read the result at 10-15 mins. Results after 20 minutes are invalid.
Result Interpretation Negative result: Only the quality control line C appears and the detection lines G and M do not show, it means that no antibody is detected and the result is negative.
Positive Result: If both the quality control line C and the detection line M appear =the Lyme disease IgM antibody is detected, and the result is positive for the IgM antibody. If both the quality control line C and the detection line G appear=the Lyme disease antibody is detected and the result is positive for the IgG antibody. If both the quality control line C and the detection lines G and M appear=the Lyme disease IgG and IgM antibodies are detected, and the result is positive for both IgG and IgM antibodies.
Invalid Result: If the quality control line C cannot be observed, results will be invalid regardless of whether a test line shows, and the test should be repeated.
Package Contents Test kit
Buffer
Dropper
Specimen transport bag
Disposable lancet
Instructions for use

For in vitro diagnostic use.