Blood Glucose Monitoring System, 500 Test Results

Blood Glucose Monitoring System, 500 Test Results

Product Introduction

The blood glucose self-monitoring system is an over-the-counter device used to quantitatively measure the glucose (sugar) content in fresh capillary whole blood samples collected from the fingertip. The blood glucose self-monitoring system is intended for use by a single person only and must not be shared with others.

How to Test System Specifications Operating Range Features Applicable Scenarios

How to Test

Control Solution 1.Gather and check supplies.
2.Allow control solution, vial of test strips and meter to adjust to room temperature for 10 minutes.
3.Wash hands. Dry thoroughly.
4.Gently swirl or invert control solution bottle to mix.
5.Remove one test strip from vial. Close vial immediately.
6.Insert test strip into Test Port. Meter turns on.
7.Remove cap from control solution bottle. Gently squeeze one drop onto a clean tissue. Wipe off bottle tip and discard tissue.
8.Gently squeeze a drop onto a small piece of unused aluminum foil, clear plastic wrap, or waxed paper . Wipe off bottle tip and discard tissue.
9.With test strip still in meter, bring Tip of test strip to top of sample drop after Drop Symbol appears in the Display. Allow drop to be drawn into test strip.
10.Remove test strip from drop when dashes appear across the meter Display. Meter is testing.
11.After testing is finished, result appears in the meter Display with the Control Symbol.
12.Compare result to Control Test Range printed on the test strip vial label for the level of control solution you are testing. If result is in range, system can be used for testing blood.
13.After result is shown, remove test strip from meter and discard. Meter turns off. Recap control solution bottle tightly.
Blood 1.Check Meter time and date before each glucose test. With meter off, press and hold the “ • ” Button until the full Display is shown and begins to blink. Release “ • ” Button. If time and date are incorrect, see Time/ Date Set Up to see the correct time and date.
2.Check supplies.
3.Allow vial of test strips and meter to adjust to room temperature for 10 minutes.
4.Clean and disinfect fingertip.
5.Remove one test strip from vial. Close test strip vial immediately.
6.Insert test strip firmly into Test Port. Meter turns on.
7.Wait until Drop Symbol appears in the Display.
8.Lance fingertip. Allow blood drop to form.
9.With test strip in Meter, bring to top of blood drop after Drop Symbol appears in the Display. Allow blood to be drawn into the test strip.
10.Remove test strip from drop when dashes appear across the meter Display. Meter is testing.
11.After testing is finished, result appears in the meter Display. Record result in log book.
12.Remove test strip from meter and discard according to your Doctor or Healthcare Professional's instructions. Meter turns off. Result is stored in the Memory.

System Specifications

Test Sample Fresh apillary whole blood from the finger or forearm, or control solution
Result Range 20-600 mg/dL
Sample Size 0.5 µL
Test Time 4 seconds
Result Value Plasma equivalent values
Assay Method Electrochemical
Power Supply One 3V lithium battery
Battery Life Approximately 1,500 tests or 2 years
Automatic Shut-off After two minutes of non-use
Weight 0.58 ounces
Size 1.6” x 1.4” x 0.85”
Memory Size 500 results

Operating Range

Relative Humidity 10% - 90% (non-condensing)
Temperature 41°F - 104°F
Hematocrit 20%-70%
Altitude Up to and including 10,200 feet
  • No programming required
  • Easy data download
  • Results in as little as 4 seconds
  • Requires only a 0.5 μL blood sample
  • Stores 500 results with date and time stamps
  • Supports calculation of 7-day, 14-day, and 30-day averages
  • Quality control testing
  • Before meal
  • After meal
  • Exercise
  • Sick
  • Medication
  • Other
Online Inquiry
0
0

There is no product in your cart.